Micomme Non-invasive Ventilator ST-30H
Description
- Non-invasive ventilation (NIV) supports the patient’s breathing without the need for intubation or a tracheotomy. NIV delivers effective therapy with less risk of infection and improved survival in patients with respiratory failure.
- Non-invasive ventilation (NIV) is the use of breathing support administered through a face mask, nasal mask, or a helmet. Air, usually with added oxygen, is given through the mask under positive pressure; generally the amount of pressure is alternated depending on whether someone is breathing in or out.
Application
- Chronic Obstruction Pulmonary Disease: The use of non-invasive ventilation (NIV) to support patients during acute decompensated respiratory failure secondary to an exacerbation of chronic obstructive pulmonary disease has unequivocal evidence of benefit in terms of reduction in the need for intubation, length of hospital stay and mortality.
- Acute respiratory distress syndrome (ARDS) is characterised by acute onset respiratory failure, diffuse pulmonary opacities and severe hypoxemia. The use of NIV can help to avoid such complications such as a higher incidence of ventilator-associated pneumonia and barotrauma.
Advantage
Max flow:210L/min
Maximum inspiratory pressure: 30cm H2O
Leakage compensation performance, the device offers 90L/min maximum leakage compensation, it can also achieve a preset target pressure to form a pressure platform to complete accurate trigger and switch.
6-level (100~600ms) pressure rise time, to meet the physiological needs of different flow rates at the beginning of inspiration.
Specification
Parameter |
ST-30H |
Ventilation mode |
S/T, CPAP, S, T, PC, VAT |
Oxygen concentration |
21%~100%, (increment by 1%) |
Screen size |
5.7 Inch color screen |
Waveform display |
Pressure/flow |
IPAP |
4~30cm H2O |
EPAP |
4~25cm H2O |
CPAP |
4~20cm H2O |
Target tidal volume |
20~2500mL |
Backup BPM |
1~60BPM |
Backup time |
0.2~4.0S |
Rise time |
1~6 level |
Ramp time |
0~60min |
Ramp pressure |
CPAP mode: 4~20cm H2O Other mode: 4~25cm H2O |
Pressure relief |
1~3 level |
Spontaneous Timin |
0.2~4.0S |
Spontaneous Timax |
0.2~4.0S |
I-Trigger setting |
Auto, 1~3 level |
E-Trigger setting |
Auto, 1~3 level |
Trigger lock |
Off, 0.3~1.5S |
Flow of HFNC mode |
N/A |
Max flow |
210L/min |
Max leak compensation |
90L/min |
Method of pressure measure |
The pressure testing tube is on mask side |
Alarms |
Apnea|Disconnection|Low minute volume|Low tidal volume|Power off|Over high pressure|Oxygen unavailable|Excessive oxygen pressure supply|Low oxygen pressure supply|Pressure tube off|Turbine dysfunction|Oxygen sensor failure|Air flow sensor failure|Low pressure|Low battery|Battery depleted |
Apnea alarm range setting |
0S, 10S, 20S, 30S |
Disconnection alarm range setting |
0S, 15S, 60S |
Real-time monitoring data |
Present oxygen concentration|Oxygen source pressure|Pressure|Ventilation per min|Respiratory rate|Current leakage|Current volume|Trigger method |
Other settings |
Screen lock|Display brightness| Flow|Pressure| Waveform |
Backup battery |
8 hours |